Metrohm 754 Dialysis Unit User Manual

Page 14

Advertising
background image

2.3 Attaching the accessories

754 Dialysis Unit

11

2 Connect new feedthrough

•

Using a wrench or a spanner, unscrew the nut on the interior

side of feedthrough 27

27 and remove the feedthrough.

•

Insert new 6.2756.000 tubing feedthrough in the hole and

retighten the nut.

3 Attach tubing to valve A

•

Screw 6.1825.210 PEEK sample loop (20

µ

L) to connections

"3" and "6" of valve A 18

18 using the enclosed 6.2744.010 PEEK

compression fittings.

•

Screw PEEK capillary 19

19 (6.1831.060; L =1 m) to connection

"2" of valve A 18

18 using a 6.2744.010 PEEK compression fit-

ting. Pull the other end of the tubing to the back through the

rear panel opening 41

41.

•

Screw PEEK capillary 20

20 (6.1831.040; L =15 cm) to connec-

tion "1" of valve A 18

18 using a 6.2744.010 PEEK compression

fitting.

4 Attach tubing to valve B

•

Screw the PEEK capillary 20

20 (6.1831.040; L =15 cm) at-

tached to connection "1" of valve A 18

18 to connection "3" of

valve B 22

22 using a 6.2744.010 PEEK compression fitting.

•

Screw PEEK capillary 21

21 (6.1831.040; L =15 cm) to connec-

tions "1" and "2" of valve B 22

22 using two 6.2744.010 PEEK

compression fittings.

•

Screw PEEK capillary 25

25 (6.1831.050; L =40 cm) to connec-

tion "4" of valve B 22

22 using a 6.2744.010 PEEK compression

fitting and pull the other end to the front through feedthrough

27

27.

•

Screw PEEK capillary 24

24 (6.1831.050; L =40 cm) to connec-

tion "5" of valve B 22

22 using a 6.2744.010 PEEK compression

fitting and pull the other end to the front through feedthrough

27

27.

•

Screw PEEK capillary 23

23 (6.1831.050; L =40 cm) to connec-

tion "6" of valve B 22

22 using a 6.2744.010 PEEK compression

fitting and pull the other end to the front through feedthrough

28

28.

5 Connecting to the 753 Suppressor Module (Option)

•

Insert the suppressor block in the 733 IC Separation Center

as per the 753 Instructions for Use.

•

Connect tubing to the IC column, the detector block and the

waste in the 733 IC Separation Center as per the 753 Instruc-

tions for Use.

•

Pull the two inlet capillaries for H

2

SO

4

and H

2

O to the front

through feedthrough 27

27 and connect to the 753 Suppressor

Module pump tubings.