4 Malfunctions – Maintenance

754 Dialysis Unit

32

4.2.6

Replacing the dialysis membrane

The dialysis membrane may have to be replaced in the following cases

(see section 4.1 as well):

•

Reduced yield from dialysis

•

Membrane has dried out, or is damaged by depos-

its or bacterial growth

•

Irremediable blockage of the sample channel (sample

can no longer be conveyed through the dialysis cell)

To replace the membrane, proceed as follows (see Fig. 7):

1 Dismantle the dialysis cell

•

Unscrew the four inlet/outlet leads from the dialysis cell

34

34

and remove cell from cell holder

66.

•

Loosen the 5 screws

39

39 completely using the 6.2621.070

Allen key and separate the top part

38

38 from the bottom part

35

35.

•

Remove old dialysis membrane

37

37.

Use only ultrapure water or ethanol to clean the dialysis cell;

organic solvents (e.g. acetone) will cause damage to the plexiglas

cell !

2 Clean dialysis cell

•

Rinse the sealing ring

36

36, the bottom part 35

35 and the top part

38

38 of the dialysis cell thoroughly with ultrapure water and dry

with N

2

or a lint-free cloth.

3 Prepare dialysis membrane

•

Take a new dialysis membrane

38

38 (e.g. 6.2714.010) out of its

packaging using the 6.2831.010 pincette and immerse in a

Petri dish containing ultrapure water for approx. 2 min until

the membrane is completely saturated with water.

4 Insert dialysis membrane

•

Place bottom part

35

35 on a paper wipe with the inside facing

upwards.

•

Insert sealing ring

36

36 in the appropriate recess in the bottom

part

35

35.

•

Using the 6.2831.010 pincette, insert the wet dialysis mem-

brane

38

38 (e.g. 6.2714.010) inside the sealing ring 36

36 on the

bottom part

35

35.