3 Operation

754 Dialysis Unit

28

3 Starting up the IC System

•

Switch on the 732 IC Detector, load method and set program

status to

active

(see the 732/733 Manual).

•

Switch on 709 IC Pump and condition IC column (see the 709

Manual).

•

If applicable, switch on the 753 Suppressor Module and the

remote control (see the 753 Manual).

•

Switch on data evaluation system (e.g. «IC Metrodata for

Win95»).

4 Starting up the dialysis system

•

Switch on the 754 Dialysis Unit and switch off the remote

control with the <REMOTE> key 22 (see section 3.1).

•

Immerse suction tubings 26

26 and 27

27 in the acceptor solution,

switch on the pump with the <PUMP> key 1

1 and rinse the

dialysis system with acceptor solution for approx. 10 min.

•

Switch off the pump with the <PUMP> key 11 and switch on

the remote control with the <REMOTE> key 22.

5 Calibration

•

Immerse sample suction tubing 26

26 in the standard solution.

•

Start program on the 732 IC Detector with the <PROG R/S>

key. The LED in the key will start to flash.

•

The next standard can be measured as soon as the LED in

the <PROG R/S> key stops flashing.

6 Measuring the sample

•

Immerse sample suction tubing 26

26 in the sample solution.

•

Start program on the 732 IC Detector with the <PROG R/S>

key. The LED in the key will start to flash.

•

The next sample can be measured as soon as the LED in the

<PROG R/S> key stops flashing.

7 Rinse the dialysis system

•

After completing the measurements, switch off the remote

control on the 754 Dialysis Unit with the <REMOTE> key 22

(see section 3.1).

•

Immerse sample suction tubing 26

26 in the acceptor solution,

switch on the pump with the <PUMP> key 1

1 and rinse the

dialysis system with acceptor solution for approx. 10 min.

•

Switch off the pump with the <PUMP> key 11.

8 Shutting down the dialysis system

•

If the dialysis cell is to remain switched off for a long period of

time, remove the inlet and outlet leads connecting to the di-

alysis cell and block openings 41

41, 42

42, 43

43 and 44

44 (see Fig. 7)

with 6.2744.060 dummy stoppers.