3 the dialysis procedure, 1 dialysis flow chart – Metrohm 754 Dialysis Unit User Manual
Page 26

3.3 The dialysis procedure
754 Dialysis Unit
23
3.3
The dialysis procedure
3.3.1
Dialysis flow chart
Every sample dialysed with the 754 Dialysis Unit comprises the follow-
ing four steps:
1 Rinsing
Valve A: FILL
Valve B: INJECT
Sample channel, ac-
ceptor channel and
sample loop are rinsed
(rinsing time: approx. 2
min).
A
B
Fill
Inject
2 Dialysis with «Stopped Flow»
Valve A: FILL
Valve B: FILL
The sample is dialysed
with resting acceptor
solution (dialysis time:
approx. 10 min).
A
B
Fill
Fill
3 Transfer to sample loop
Valve A: FILL
Valve B: INJECT
The acceptor solution
enriched with ions from
the sample is trans-
ferred to the sample
loop (transfer time:
approx. 0.4 to 0.6 min).
A
B
Fill
Inject
4 Injection
Valve A: INJECT
Valve B: INJECT
The acceptor solution
enriched with ions from
the sample is injected
into the IC system.
A
B
Inject
Inject
Eluent
Column
Waste
Sample
Acceptor
Eluent
Column
Waste
Sample
Acceptor
Eluent
Column
Waste
Sample
Acceptor
Eluent
Column
Waste
Sample
Acceptor