3.3 The dialysis procedure

754 Dialysis Unit

27

2 Measure the standard with dialysis

•

Set the dialysis time in the program to 5 min (see section

3.3.2).

•

Immerse sample suction tubing 26

26 in the standard used in

point 1.

•

Start the program on the 732 IC Detector with the <PROG

R/S> key. Wait until the LED in the <PROG R/S> key stops

flashing, signalling the end of the evaluation.

3 Additional measurements

•

Increase the dialysis time in the program (see section 3.3.2)

by 5 min increments until the measured values remain con-

stant.

•

Take measurements as per point 2.

4 Determine optimum dialysis time

•

Record the ratio of the peak surfaces, i.e. concentrations

measured with and without dialysis in relation to the dialysis

time (i.e. the recovery rate) and determine the optimum dialy-

sis time.

3.3.6

Setting the program time

Set the program time so that the chromatogram can be completely re-

corded and evaluated within this period.

3.3.7

Dialysis procedure

To analyse samples using dialysis you should preferably proceed as

follows:

1 Prepare acceptor solution

•

To avoid malfunctions caused by air bubbles in the acceptor

channel, the ultrapure water used for the acceptor solution

should always be degassed for at least 10 min with vacuum,

N

2

or He.

2 Prepare sample

•

To prevent blockages in the sample channel, samples con-

taining a high volume of (suspended/solid particles) should

always be centrifuged using a table centrifuge at 10,000

rev./min for 5 min.