14 Program settings

TiNet 2.5 Compliant

Enforce login with password ([on], off)

If this option is enabled, every user must enter user name and
password in the

Login

window at every program start. If this op-

tion is disabled, no login is required. The following login and
password policies are meaningless and will be set to inactive
(gray).

Maximum number of login attempts ([on], off); 2...[3]...6)

If this option is enabled, a user is set to "

inactive

" as soon as the

maximum number of login attempts is exceeded. The counter
for login attempts is set to zero for all users at every new start
of the program.

Automatic logout after ([on], off); 1...[10]...60 min)

If this option is enabled, the user is logged out automatically if
no operating function via mouse or keyboard is done during
the defined delay time.

New login only for the same user ([on], off)

If this option is enabled, only the same user can login after
automatic logout. In any case, user with administrator privi-
leges can always login.

Password

Enforce unique password ([on], off)

If this option is enabled, passwords must be unique. Expired
passwords are treated like active passwords, so every pass-
word can only be used once.

Enforce use of special characters ([on], off)

If this option is enabled, a password must contain at least one
special character.

Minimum password length ([on], off; 1...[6]...40 characters)

If this option is enabled, the password must have the set num-
ber of characters.

Passwords expire every ([on], off; 1...[365]...999 days)

If this option is enabled, a new password must be entered after
the expiration time.

Advertising