5 sign result – Metrohm TiNet 2.5 User Manual
Page 59

TiNet 2.5 Compliant
Results • 57
This report contains the following entries (independent of the set column
view):
Date:
Date when determination was started.
Time:
Time when determination was started.
User:
Name of the user who has started the determination.
Remark:
Remark entered on the
Sample data
tab.
Method:
Method
name.
Version:
Version number of the method
Id1...Id7
Sample identifications (all identifications which are not empty
are printed).
Sample size
Sample size (mass).
Unit
Unit of the sample size.
Results
Results (all results which are not empty are displayed)
Report printed on:
Date and time when the result report was printed.
Signed by:
Full name of the user who signed the results.
Signed on:
Date and time when the results were signed.
Sign comment:
Comment for the signature.
6.3.5 Sign
result
Results overview, File, Sign result
With this menu item, the
Sign-off
window appears for applying electronic
signatures on results of the selected determination (the old signature data
is overwritten).