4.2 Diagnosis
784 KFP Titrino
107
8.1 RS232 test 1
•
Press
<1>..
RS232 test-connector? 1
•
Insert the 3.496.8560 test plug in port ‘A1’.
•
Press
<ENTER>
.
Test sequence:
1. In case of an error an error message is displayed. If for example no test plug is
connected, error message error 69 appears).
2. If no error is found, after about 5 s display shows
RS232 test 1 o.k.
•
Remove test plug.
•
Press
<CLEAR>.
RS232 test 1...2
8.2 RS232 test 2
•
Press <2>.
RS232 test-connector? 2
•
Insert the 3.496.8560 test plug in port ‘A2’.
•
Press
<ENTER>.
Test sequence:
1. In case of an error an error message is displayed. If for example no test plug is
connected, error message error 69 appears)
2. If no error is found, after about 5 s display shows
RS232 test 2 o.k.
•
Remove test plug.
•
Press
<CLEAR>.
RS232 test 1...2
•
Press <CLEAR>.
diagnose press key 0...9