3.2 Remote control commands

784 KFP Titrino

89

.Feed:

Length of the time axis. Depending on the printer, the measure in
cm may not always be correct.

Curve length

0.01

100 cm

0.1

10 cm

0.5

2 cm

1

1 cm

3.2.2.94. Setup.PowerOn

$G

Simulation of 'power on'. The device has the same status as after power on:
The cylinder is filled, error messages deleted and the current sample number
set to 0. The method last used is ready for operation.

3.2.2.95. Setup.Initialise

$G

Setup.Initialise.Select

ActMeth, Silo, Config,

Assembly, Setup, All

Setting of default values for the following areas:
ActMeth: Current method. Parameters, calculations, and assignments for

the data output, operands C01...C19.

Silo:

The silo memory is deleted. Same function as delete entire silo.

Config:

All values under &Config.

Assembly: All values under &Assembly.
Setup:

All values under &Setup.

All:

Values of the entire tree (except silo and method memory).

The action must be triggered with &Setup.Initalise $G.

3.2.2.96. Setup.RamInit

$G

Initializes instrument, see page 110. All parameters are set to their default
value and error messages are cleared. The user and silo memories will be
deleted. The user memory contains the default user methods from Metrohm.

3.2.2.97. Setup.InstrNo

$G

Setup.InstrNo.Value

serial number, 8 ASCII characters

Instrument identification for report output.
Set the value with &Setup.InstrNo $G .

3.2.2.98. Diagnose.Report

$G

Output of the report containing the adjustment parameters. The Titrino has to
be in its inactive basic state.