2.6 Parameters, key <PARAM>

784 KFP Titrino

24

In the following table we attempt to show you solutions related more to the instrument's

side:

What to do if ...

Problem

Possible causes and remedial action

Dosing at end too long and

increments too small. "Is never

finished!"

•

Increase "min.volume incr.".

•

Change stop criterion. Try to increase stop drift or

use a short stopping time as stop criterion, e.g.

•

For problematic samples change solvents with

ketones or aldehydes in 2-methoxyethanol or with

amines mixture in methanol/glacial acetic, e.g., see

literature.

The increments at the end of the

titration are too large.

"Overshoots".

•

Lower "max.rate". The following experiment gives you

a reference point for the optimum max.rate: Drift

display during conditioning and add sample without

starting the titration. Select a value below the

maximum drift as "max.rate".

•

Optimize setup of electrode and buret tip and

improve stirring.

Solution becomes too brown at

the end of the titration.

•

The methanol fraction in the solvent is too low.

Change the solvent.

•

Electrode could be coated; wipe off with acetone.

Solution becomes darker with

every titration.

•

Renew solvent.

•

Electrode could be coated; wipe off with acetone.

The drift increases with every

titration.

•

Does your sample evolve water extremely sluggishly?

Work with the KF oven.

•

Are acids esterified in your sample? Change solution

more frequently. Increase buffer capacity of the

solvent.

•

Does your sample contain ketones or aldehydes?

Use special reagents suitable for ketones and

aldehydes.

The endpoint is reached "too

rapidly".

•

Reduce "max.rate".

The titration times become

longer and longer.

•

With 2-component reagents the buffer capacity of the

solvent can be exhausted. Change solution.

•

If the drift increases at the same time, see above.