Metrohm 784 KFP Titrino User Manual
Page 27

2.6 Parameters, key <PARAM>
784 KFP Titrino
24
In the following table we attempt to show you solutions related more to the instrument's
side:
What to do if ...
Problem
Possible causes and remedial action
Dosing at end too long and
increments too small. "Is never
finished!"
•
Increase "min.volume incr.".
•
Change stop criterion. Try to increase stop drift or
use a short stopping time as stop criterion, e.g.
•
For problematic samples change solvents with
ketones or aldehydes in 2-methoxyethanol or with
amines mixture in methanol/glacial acetic, e.g., see
literature.
The increments at the end of the
titration are too large.
"Overshoots".
•
Lower "max.rate". The following experiment gives you
a reference point for the optimum max.rate: Drift
display during conditioning and add sample without
starting the titration. Select a value below the
maximum drift as "max.rate".
•
Optimize setup of electrode and buret tip and
improve stirring.
Solution becomes too brown at
the end of the titration.
•
The methanol fraction in the solvent is too low.
Change the solvent.
•
Electrode could be coated; wipe off with acetone.
Solution becomes darker with
every titration.
•
Renew solvent.
•
Electrode could be coated; wipe off with acetone.
The drift increases with every
titration.
•
Does your sample evolve water extremely sluggishly?
Work with the KF oven.
•
Are acids esterified in your sample? Change solution
more frequently. Increase buffer capacity of the
solvent.
•
Does your sample contain ketones or aldehydes?
Use special reagents suitable for ketones and
aldehydes.
The endpoint is reached "too
rapidly".
•
Reduce "max.rate".
The titration times become
longer and longer.
•
With 2-component reagents the buffer capacity of the
solvent can be exhausted. Change solution.
•
If the drift increases at the same time, see above.