2.6 Parameters, key <PARAM>

784 KFP Titrino

22

Mechanisms of the KFT parameters in Ipol mode

The position and curve characteristics of the line between the ranges of excess water

or excess iodine depends on the type of sample and the ingredients of the working

media.

The endpoint has to be set close to the range margin, but always within the iodine

excess range. If the endpoint is set too close to the limit, an overdose of KF reagent

may be dispensed. The steeper the curve at the preset polarization current I(pol), the

more delicate it is to titrate to a steady and reproducible endpoint. For the titration of

troublesome samples, the particular conditions must be optimized by trial and error.

Always mind the color of the working medium at the end point as your guideline.

Remark: In most cases the preset default parameters provide precise and

reproducible results.

If you select negative values for a parameter, you should choose negative values for all

other parameters too in order to have compatible parameter sets.

In KFT Upol mode, all parameters work in a similar manner as in KFT Ipol mode.