Metrohm 784 KFP Titrino User Manual
Page 129

6.2 Pin assignment of the "Remote" socket
784 KFP Titrino
126
Outputs
pin 17 (Output 3)
pin 3 (Output 4)
pin 16 (Output 5)
pin 1 (Output 6)
pin 2 (Output 7)
pin 6 (Output 8)
pin 7 (Output 9)
pin 8 (Output 10)
pin 13 (Output 11)
pin 19 (Output 12)
pin 20 (Output 13)
End of determination
EOD
not used
Error,
active with errors
Activate pulse, see page 128.
L6 in TIP
Pulses for recorder (t
p
=150
µ
s)
10 000 per buret cylinder
not used
not used
not used
not used
sample size out of limits
result out of limits, see page 26
For all outputs:
V
CE0
= 40 V
I
C
= 20 mA
t
Pulse
> 100 ms
Functions see page 127.
Voltage
pin 15
pin 14
pin 25
I
≤
200 mA
0 V: active
5 V: inactive
Ordering numbers for plug:
K.210.9004 (shell) and K.210.002
No liability whatsoever will be accepted for damage caused by im-
proper interconnection of instruments.
0 V
+5 V