Metrohm 784 KFP Titrino User Manual
Page 30

2.7 Result calculations
784 KFP Titrino
27
Input of method specific operands C01...C19, key <C-FMLA>
With <C-FMLA> the operands C01...C19 can be put
in. For the calculation, the operands are used, which
were introduced in the formula.
The inputs method specific and are store in method
memory.
The calculation report can be printed with the key sequence
<PRINT><
←
/
→
> (press keys repeatedly until "calc" appears in the display) <ENTER>.
Example for a calculation report:
Calculation report
mode und method name
formulas:
result name=formula;number of decimal
places;result unit
sample size
method specific operands
values of the common variables that are
used
C-FMLA
C
1
'fm
784 KFP Titrino 02134 784.0010
user sn
date 1999-08-15 time 15:01 3
KFT Ipol KF-Blank
>calculations
Water=(EP1-C38)*C39*C01/C00/C02;2;%
Titer=C39;4;mg/ml
Blank=C38;4;ml
C00= 1.0
C01= 0.1
C02= 1
C38= 0.0
C39= 0.0
------------