7.2 Monotonic equivalence point titrations (MET)
■■■■■■■■■■■■■■■■■■■■■■
108
■■■■■■■■
862 Compact Titrosampler
Measuring mode Upol:
Range
–125.00 ... 125.00 µA
Default value
–125.00 µA
Upper limit
This parameter is visible only when Window = on.
Measured value for the upper limit.
Measuring mode pH:
Range
–20.000 ... 20.000
Default value
20.000
Measuring mode U, Ipol:
Range
–1250.0 ... 1250.0 mV
Default value
1250.0 mV
Measuring mode Upol:
Range
–125.00 ... 125.00 µA
Default value
125.00 µA
EP criterion
The equivalence point criterion found (ERC = Equivalence point Recogni-
tion Criterion) is compared with this value. Equivalence points whose ERC
is less than the value defined here will be ignored.
Measuring mode pH:
Range
0.10 ... 9.99
Default value
0.50
Measuring mode U, Ipol:
Range
1 ... 999 mV
Default value
30 mV
Measuring mode Upol:
Range
0.1 ... 99.9 µA
Default value
2.0 µA
EP recognition
This parameter allows you to filter out only the equivalence points that are
being sought.
for Window = off
Selection
all | greatest | last | off
Default value
all