7 Parameters

862 Compact Titrosampler

■■■■■■■■

133

Note

The setting of the stirring rate under Menu

▶ Parameters ▶ Automa-

tion applies only for stirring during the immersion of the electrode in
the rinsing beaker. The stirring rate during the determination is set
under Menu

▶ Parameters ▶ Titration parameters.

7.8

Automation: Rinsing in sample

Menu

▶ Parameters ▶ Automation

Automation

Display of the template used for the automation sequence.

Dripping time

Waiting time after the titration head moves out of the sample beaker.

Range

0 ... 999 s

Default value

3 s

Aspiration time

Aspiration time of Pump 2, in case an 843 Pump Station is connected. It
runs before the rinsing and after the rinsing.

Range

0 ... 999 s

Default value

10 s

Rinsing time

Rinsing time of Pump 1, if an 843 Pump Station is connected. Rinsing
pump and aspiration pump run during the rinsing time.

Range

0 ... 999 s

Default value

5 s