7 Parameters
862 Compact Titrosampler
■■■■■■■■
133
Note
The setting of the stirring rate under Menu
▶ Parameters ▶ Automa-
tion applies only for stirring during the immersion of the electrode in
the rinsing beaker. The stirring rate during the determination is set
under Menu
▶ Parameters ▶ Titration parameters.
7.8
Automation: Rinsing in sample
Menu
▶ Parameters ▶ Automation
Automation
Display of the template used for the automation sequence.
Dripping time
Waiting time after the titration head moves out of the sample beaker.
Range
0 ... 999 s
Default value
3 s
Aspiration time
Aspiration time of Pump 2, in case an 843 Pump Station is connected. It
runs before the rinsing and after the rinsing.
Range
0 ... 999 s
Default value
10 s
Rinsing time
Rinsing time of Pump 1, if an 843 Pump Station is connected. Rinsing
pump and aspiration pump run during the rinsing time.
Range
0 ... 999 s
Default value
5 s