2 entering sample data in the main dialog, Entering sample data in the main dialog – Metrohm 862 Compact Titrosampler User Manual
Page 58

5.5 Sample data
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
862 Compact Titrosampler
5.5.2
Entering sample data in the main dialog
In the main dialog you can enter the sample data even while a determina-
tion is running (see Chapter 5.7, page 53). It will be used for the run-
ning determination.
ID1
Sample identification. The sample identification can be used in calculations
as the variable CI1.
Input
max. 10 characters
Default value
empty
ID2
Sample identification. The sample identification can be used in calculations
as the variable CI2.
Input
max. 10 characters
Default value
empty
Sample size
Sample size. The value of the sample size can be used in calculations as
the variable C00.
Range
–999999999 ... 9999999999
Default value
1.0
Unit
Unit of sample size.
Selection
g | mg | µg | mL | µL | pieces | User-defined
Default value
g
User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined.