Remote interface
10.6.1
Pin assignment of the remote interface
1
5
6
9
5
1
9
6
Figure 32
Pin assignment of remote socket and plug
The above figure of the pin assignment applies for all Metrohm instru-
ments with 9-pin D-Sub remote connectors.
Table 7
Inputs and outputs of the remote interface
Pin No.
Assigment
Function
1
Output 0
Sample ready
2
Output 1
Dosimat
3
Output 2
Pump 1
4
Output 3
Pump 2
5
Output 4
Error
6
0 Volt (GND)
7
+5 volts
8
Input 0
Start
9
Input 1
Stop
Inputs
+5 V
approx. 5 k
Ω Pull-up
tp
t
p
> 100 ms
active = low, inactive = high