7 Parameters
862 Compact Titrosampler
■■■■■■■■
129
Number of buffers
Number of buffers that are used for calibration. If calibration is accom-
plished with more than two buffers, then they can be used repeatedly
give them more statistical weight. The first two buffers must however
always be different from one another.
Selection
1 | 2 | 3 | 4 | 5
Default value
2
Buffer 1 pH
This parameter is visible only when Buffer type = Special.
Range
–20.000 ... 20.000
Default value
7.000
Buffer 2 pH
This parameter is visible only when Buffer type = Special.
Range
–20.000 ... 20.000
Default value
4.000
Selection
off
Buffer 3 pH
This parameter is visible only when Buffer type = Special.
Range
–20.000 ... 20.000
Selection
off
Default value
off
Buffer 4 pH
See Buffer 3 pH.
Buffer 5 pH
See Buffer 3 pH.
7.4.3
Reports
Menu
▶ Parameters ▶ Reports
The reports that will be printed out automatically in connection with a cal-
ibration are defined under Reports.
Results
The result report contains the specifications for the calibration (slope,
pH(0), etc.).
Selection
on | off
Default value
off