9 automation: rinsing in special, Automation: rinsing in special – Metrohm 862 Compact Titrosampler User Manual
Page 144

7.9 Automation: Rinsing in special
■■■■■■■■■■■■■■■■■■■■■■
134
■■■■■■■■
862 Compact Titrosampler
7.9
Automation: Rinsing in special
Menu
▶ Parameters ▶ Automation
Automation
Display of the template used for the automation sequence.
Dripping time
Waiting time after the titration head moves out of the sample beaker and
out of the rinsing beaker.
Range
0 ... 999 s
Default value
3 s
Aspiration time
Aspiration time of Pump 2, in case an 843 Pump Station is connected. It
runs after the rinsing time.
Range
0 ... 999 s
Default value
10 s
Rinsing time
Rinsing time of Pump 1, in case an 843 Pump Station is connected. It runs
before the aspiration time. Rinsing pump and aspiration pump run during
the rinsing time.
At the end of a sample series, the rinsing time determines how long rins-
ing solution will be filled into the rinsing beaker.
Range
0 ... 999 s
Default value
5 s