11. Introduction
. Introduction
726 Titroprocessor, Instructions for use
55
1.3.5 Measuring interfaces / Measuring inputs
The 726 Titroprocessor models 2.726.0110 and 2.726.0120 are equipped
with 2 measuring interfaces (Sensors A and Sensors B).
The diagram below shows where the sensors are to be connected to meas-
uring interface A, which is found on all 726 Titroprocessor models.
1
Ind A2 Ind A1
Ref
Pol
Pt100/1000
Sensors A
1.3.6 Data cards
Data cards can be used to store methods or measuring data and to transfer
them from one instrument to another. They are an excellent mobile storage
medium.
Data
card
Attention must be paid to several points when handling data cards.
So-called PC cards which meet the Standard PCMCIA 2.x (68 Pins) / JEIDA
4.x can be used.
Flash cards
can be read but not written on, e.g. the application card
(Order. no. 6.6024.000) supplied with the Titroprocessor
Stirrer Int A con-
nector
Indicator electrode
A2
Indicator electrode
A1
Pt100 / Pt1000
Temperature sensor
Reference
electrode
Polarized
electrode
Right-hand side of housing
Eject button
Notch