8. Command reference
8. Command reference
726 Titroprocessor, Instructions for use
157
157
tion and to output it as a temperature curve.
numerical
manual input (e.g. for pH correction)
auto
automatic measurement during the
whole titration (for temperature curve)
initial
one-off measurement at the start of the
titration mode
The temperature can be included in formulas for calcula-
tions (see page 118ff), e.g. for volume corrections. The dif-
ferent system variables are listed on page 148 and 128.
Please note: start temperature ST and closing tempera-
ture CT are only available with auto.
U(pol)
Polarization voltage (only with SEC_UPOL*)
-1270…400…1270 mV
The polarization voltage is the voltage which is applied to a
polarized electrode during amperometric titration. The re-
sultant measuring quantity is µA.
I(pol)
Polarization current (only with SEC_IPOL*)
-127…1…127 µA
The polarization current is the current which is applied to a
polarized electrode during voltametric titration. The resul-
tant measuring quantity is mV.