Metrohm 726 Titroprocessor User Manual
Page 235
11. File manager
726 Titroprocessor, Instructions for use
229
229
•
If problems occur during data reception it may be necessary to
deactivate the FIFO buffer of the RS232 interface in the system settings
of the computer or to reduce the baud rate of both instruments to 2400
Baud. Consult the instructions for use of the Metrodata 726 backup
software.
In order to start data transmission the ‘Receive’ or ‘Send’ mode of the
backup software at the PC must be switched on before the [Send to RS]
or [Receive from RS] softkey of the Titroprocessor is pressed. After
selection of the RS232 interface transmission takes place.
Files (method, silo data, determination files) can also be transmitted or
received directly from the file dialog window of the corresponding dialog
page. Select 'Storage' 'RS232 x' and press the [Load] or [Save]
softkey.