2.3 Mains connection

726 Titroprocessor, Instructions for use

16

16

WARNING - Fire Hazard -

with the same type and rating of fuse

For continued protection replace only

50-60Hz 160VA

2A(TH)

115V

230V

230V

115V: 100V...120V 10%
230V: 220V...240V 10%

2.3 Mains connection

2.3.1 Fuses

Two fuses

S1

S1

and

S2

S2

are built into the 726 Titroprocessor as standard;

both are type 2 ATH (2 A, slow-blow, with high switching capacity, Metrohm

order no. U.600.0107).

Make sure that the instrument is never operated with any different type
of fuse as otherwise there is a fire hazard!

To change blown fuses proceed as follows:

Exchange fuse S1 (zero conductor, 2 ATH)

•

Remove mains cable from mains connector.

•

Use a screwdriver to turn fuse holder

S1

S1

to the left until it can be pulled

out.

•

Pull out the fuse holder, remove the blown fuse and replace it with a

spare fuse (2 ATH).

•

Insert the fuse holder in the instrument, press it down with a screwdriver

and then turn it to the right to fix it in position.

Exchange fuse S2 (phase, 2 ATH)

•

Remove mains cable from mains connector.

•

Use a screwdriver to push fuse holder

S2

S2

upwards until it can be pulled

out.

•

Pull out the fuse holder, remove the blown fuse and replace it with a

spare fuse (2 ATH).

•

Insert the fuse holder in the instrument and push until it clicks into place.

Fuse holder

Mains voltage selector

Earth socket

Mains switch

Rear view

S1

S2

Mains connection