8.13 Communication functions

726 Titroprocessor, Instructions for use

200

200

The assignment of the logical input and output lines on the pin occupancy

of the particular remote sockets is largely unified for Metrohm instruments,

so that with the help of a 6.2141.020 standard remote cable remote con-

nections can be made relatively simply and communication commands

used without any problems. If instruments from other manufacturers are

used then the technical data given in their instruction manuals should be

consulted. If the differing pin occupancies of the remote interfaces cause

any problems then the service department of your local Metrohm agency

will be able provide a suitable connection cable.

Tip: the current states of the input lines (Input 0…7) can be monitored 'live'

for the creation of a method. It is also possible to alter the output lines

(Output 0…8, or …13) manually to check their effects.

This is done by opening the dialog page 'Devices & Manual Con-
trol'

with the <MAN CONTROL> key. The control window of the remote

interface can be found under the softkey [Remote lines] (or [Sample
changer] [Remote lines]

).

With the help of the remote control lines items such as external pumps, Ion

meters or a KF Oven can be connected to the 726 Titroprocessor and con-

trolled by it. With the aid of the Metrohm 731 Relay box any other external

instruments can be controlled via remote lines.

SCAN_RM

Scanning the input lines of the remote socket

The method run is stopped until the status of the input lines corresponds to

the preset bit pattern. Short signal impulses (>200 ms) are recognized. The

statuses of the input lines are not altered by the SCAN command.

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

SCAN_RM

Remote A Remote ********

The input lines of the Titroprocessor itself as well as those of the remote

sockets of the two connected 717 Sample changers can be scanned.

Remote

Remote socket address

A | 1 |2

The Titroprocessor remote socket has the address A; the

remote socket of a sample changer is always 1 or 2, de-

pending on the EBus address of the changer.

Pattern

Bit pattern of line status

8 characters (0 | 1 | *)

An 8-place bit pattern is defined in which each bit is as-

signed to an input line. If they coincide the method run is

continued with the next command line.