3.2
Settings
824 Easy Sample Changer,
Operation
23
3.2.3 Setting the shifting position of the lift
STIR 1
STIR 2
The lift with the titration head can be easily moved up and down with the
<Ï> and <Ð> keys respectively.
Set the shift height of the lift so that the sample rack can rotate freely.
The electrode must fully emerge from the sample beaker.
Accept/ continue with <Î>.
3.2.4 Setting the stirrer rate
STIR 1
STIR 2
The stirrer rate (applies to both stirrer outputs) can be easily increased
or reduced in 15 steps with the <Ï> and <Ð> keys respectively.
The standard setting is step 3.
Accept/ continue with <Î>.
1
2
3
4
1
2
3
4