8.2 Method listings
824 Easy Sample Changer,
Annex
43
8.2.2 Method 2
824 Easy Sample Chan.
5.824.0010
¬
Report header with program version
Parameters
method
2
¬
Method name
number of samples:
rack
¬
No. of samples (whole sample rack)
>start sequence
1 CTL:Rm:
INIT
¬
Initialize remote interface
2 SAMPLE:
-
1
¬
Initialize the
3 SAMPLE:
+
1
sample position
>sample sequence
1 SHIFTRATE:
+
20
¬
Rack direction of rotation (clockwise)
2 MOVE
1
:
sample
¬
Move next sample in front of tower
3 LIFT: 1
:
work mm
¬
Move lift to working position
4 STIR: *
:
ON s
¬
Switch on stirrer
5 CTL:Rm:
START dos*
¬
Start addition
6 WAIT
10 s
¬
Waiting time
7 CTL:Rm:
START device1
¬
Start instrument/titrator
8 SCN:Rm
:
End1
¬
Wait for end of determination [EOD]
9 STIR: *
:
OFF s
¬
Switch off stirrer
10 LIFT: 1
:
shift mm
¬
Move lift to rotating position
11 WAIT
3 s
¬
Waiting time: 3 s for drips
12 SHIFTRATE:
-
20
¬
Rack direction of rotation (counterclockwise)
13 MOVE
1
:
spec.1
¬
Move conditioning beaker in front of tower
14 LIFT: 1
:
work mm
¬
Move lift to working position
15 STIR: *
:
ON s
¬
Switch on stirrer
16 WAIT
5 s
¬
Waiting time
17 STIR: *
:
OFF s
¬
Switch off stirrer
18 LIFT: 1
:
shift mm
¬
Move lift to rotating position
19 WAIT
3 s
¬
Waiting time: 3 s for drips
>final sequence
1 MOVE
1
:
spec.1
¬
Move conditioning beaker in front of tower
2 LIFT: 1
:
work mm
¬
Immerse electrode
>changer settings
------ Sample changer settings --------
rack number
0
lift rate 1
25 mm/s
shift rate
20
shift direction:
auto.
beaker test mode:
single
¬
If sample beaker is missing then the
on beaker error:
MOVE
next one will be moved to automatically
>stirring rates
------ Stirring rates ---------------------------
stirrer 1
3
stirrer 2
3
>manual stop
------ Reaction to manual stop --------------
CTL Rmt:
STOP device1
¬
Stop instrument/titrator
CTL RS232: