6 method 4, Method 4 – Metrohm 824 Easy Sample Changer User Manual

Page 39

Advertising
background image

4.6 Method 4

824 Easy Sample Changer,

Standard methods

33

4.6 Method 4

Use:

For simple titrations (including KF titrations) or pH measurements (with-

out calibration). The initial addition of auxiliary reagents or solvents is

possible.

Required instruments:

•

Metrohm titrator (Titrino or Titrando) or pH meter

•

Metrohm Dosimat (optional)

•

Metrohm 722/802 Rod Stirrer or 741 Magnetic Stirrer

•

2 external pumps (e.g. Metrohm 772 Pump Unit; Pump 1 for rins-

ing, Pump 2 for aspirating

Features:

After each determination the sample vessel is emptied by Pump 1 and

then the electrode is rinsed by Pump 2. The sample vessel is then emp-

tied again by Pump 1.

Preparation:

Fill a sufficiently large conditioning vessel with a suitable solvent and

place it in the highest rack position.

Procedure:

•

(Start sequence) Initialize the sample changer

•

Move to sample and lower lift to working position

•

Switch on stirrer

•

If a Dosimat is connected (optional): add auxiliary reagent or sol-

vent and wait until addition is completed

•

Start titration/measurement and wait until it is completed

•

Switch off stirrer

•

Move lift to rinsing position and empty sample vessel

•

Rinse electrode and empty sample vessel again

•

Move lift to rotating position and allow electrode to drip

•

(Final sequence) Move to conditioning beaker and immerse elec-

trode