LaMotte 1200 Colorimeter User Manual

Page 19

Advertising
background image

The 1200 can be calibrated with two prepared standards. The concentrations of
the standards should be chosen from the low and high ends of the range of the
meter. The low standard must be less than 5% of over range. To calibrate the
meter with two standards, follow steps 1-11 of the calibration procedure for one
reacted sample. Repeat steps 1-11 of the calibration procedure with the second
reacted standard.

The calibration procedure should be followed as often as required by
regulations and laws for compliance monitoring.

To reset the calibration to the factory calibration: with the meter off, hold
down the

t and push READ. The meter will turn on and the calibration

will be reset.

19

Push the
button again to

memorize the

calibration. The

1200 display will

stop flashing.

Calibration is

complete.

ZERO

Turn the unit off by
holding the
button down for at

least 2 seconds, or

proceed to measure

the test samples

following the

procedure on page

20.

READ

9.

10.

11.

Push and hold the

button for 5

seconds until

is

displayed. Release

the button. The

display will flash.

Adjust the display

with the

and

buttons until the

value of the standard

is displayed.

ZERO

CAL

I200

••••••••••••••••••

ZERO

COLORIMETER

READ

““

““

2.55

I200

••••••••••••••••••

ZERO

COLORIMETER

READ

2.50

I200

••••••••••••••••••

ZERO

COLORIMETER

READ