PHOSPHATE PROCEDURE - ASCORBIC ACID METHOD

Read the 1200 Colorimeter Manual before proceeding. Carefully
wipe tubes dry before inserting into the colorimeter chamber.

PHOSPHATE

4.

Push the

button

to turn the meter

on. Press the

button

and hold it for 2
seconds until

is displayed.

Release the button

to take a blank

reading (0.0 ppm).

READ

ZERO

bla

I200

COLORIMETER

••••••••••••••••••

READ

ZERO

bla

Insert the tube

into the

chamber, being

sure to align

the index line

with the arrow

on the meter.

Close the lid.

This tube is

the sample

blank or zero.

3.

Rinse and fill a

colorimeter tube

(0290) to the
10 mL line with

sample water. Cap

and wipe dry.

2.

Fill the Water

Sample Collecting

Bottle (0688) with

sample water. This

will be used to

dispense sample

water for the tests.

1.

Remove tube

from colorimeter.

Use 1.0 mL pipet

(0354) to add
1.0 mL of

*Phosphate Acid

Reagent
(V-6282). Cap

and mix.

5.

6.

Use the 0.1 g spoon

(0699) to add one

measure of *Phosphate

Reducing Reagent

(V-6283).

8.

Align the index line with the

arrow on the meter, insert tube

CAL

I200

COLORIMETER

••••••••••••••••••

READ

into chamber. Close

the lid. Push the

button.

Record results as

ppm

Orthophosphate.

READ

7.

Cap and shake until

powder dissolves.

Wait 5 minutes for

full color

development.

Solution will turn

blue if phosphates

are present. Wipe

tube dry.