CHLORINE DIOXIDE TEST PROCEDURE - DPD METHOD

Read the 1200 Colorimeter Manual before proceeding.
Carefully wipe tubes dry before inserting into the colorimeter
chamber.

CHLORINE DIOXIDE

2

Remove tube.

Add 5 drops

Glycine

Solution

(6811). Cap

and mix.

Cap and invert to

mix. Wipe tube dry.

Make readings

within 30 seconds.

Fill the Water

Sample Collecting

Bottle (0688) with

sample water. This

will be used to

dispense sample

water for the tests.

Insert the tube

into the

chamber, being

sure to align the

index line with

the arrow on

the meter.

Close the lid.

This tube is the

blank or zero.

Rinse and fill a

colorimeter tube

(0290) to the

10 mL line with

sample water.

Cap and wipe

dry.

5.

7.

3.

1.

2.

4.

6.

I200

COLORIMETER

••••••••••••••••••

READ

bla

ZERO

Add 5 drops DPD

#1A Free Chlorine

Reagent (6740),

and 5 drops of

*DPD #1B Free

Chlorine Reagent

(6741).

Push the
button to turn the

meter on. Press the

button

and hold it for 2
seconds until

is

displayed. Release

the button to take a

zero reading
(0.00 ppm).

READ

ZERO

bla

8.

CAL

I200

COLORIMETER

••••••••••••••••••

READ

Align the index

line with the arrow

on the meter, insert

tube into chamber.

Close the lid. Push

the

button. Record

results as ppm

Chlorine Dioxide.