FLUORIDE SPADNS TEST METHOD SPECIFICATIONS

APPLICATION

Drinking and surface waters; domestic and industrial waters.

RANGE

0.0 to 2.0 ppm Fluoride

METHOD

Colorimetric test based upon the reaction between fluoride and zirconium dye lake. The
fluoride reacts with the dye lake, dissociating a portion of it into a colorless complex ion
and dye. As the fluoride concentration increases, the color produced becomes
progressively lighter.

HANDLING & PRESERVATION

Samples may be stored and refrigerated in plastic containers.

INTERFERENCES

The following substances produce a positive interference at the concentration given:

Chloride (Cl

-

)

7000 mg/L

Phosphate (PO

4

-3

)

16 mg/L

Hexametaphosphate (NaPO

3

)

6

1 mg/L

The following substances produce a negative interference at the concentration given:

Alkalinity (CaCO

3

)

5000 mg/L

Aluminum (Al

+3

)

0.1 mg/L

Iron (Fe

+3

)

10 mg/L

Sulfate (SO

4

-2

)

200 mg/L

Color and turbidity must be removed or compensated for in the procedure. Temperature
should be maintained within 5 degrees Celcius of room temperature.

LaMOTTE COMPANY

Helping People Solve Analytical Challenges

®

PO Box 329 • Chestertown • Maryland • 21620 • USA

800-344-3100 • 410-778-3100 (Outside USA) • Fax 410-778-6394

Visit us on the web at www.lamotte.com