CHLORINE TEST PROCEDURE - DPD METHOD

Read the 1200 Colorimeter Manual before proceeding. Carefully
wipe tubes dry before inserting into the colorimeter chamber.

FREE AVAILABLE CHLORINE

7.

8.

5.

3.

1.

2.

4.

6.

I200

COLORIMETER

••••••••••••••••••

READ

READ

ZERO

CAL

••••••••••••••••••

bla

COLORIMETER

I200

Fill the Water

Sample Collecting

Bottle (0688) with

sample water. This

will be used to

dispense sample

water for the tests.

Insert the tube

into the

chamber, being

sure to align the

index line with

the arrow on

the meter.

Close the lid.

This tube is the

blank or zero.

Empty all but a few

drops of sample from

tube. Add one

*Chlorine DPD #1

IG Tablet (6903A) and

crush with tablet

crusher (0175).

NOTE: To insure

accurate results, tablet

must be crushed before

filling tube.

Align the index line with the

arrow on the meter, insert tube

into chamber.

the lid. Push the

READ

button.

Record results as

ppm Free Available

Chlorine. Save

sample for the Total

Residual Chlorine

test.

Rinse and fill a

colorimeter tube

(0290) to the

10 mL line with

sample water.

Cap and wipe

dry.

Push the

READ

button to turn the

meter on. Press the

ZERO

button and

hold it for 2 seconds

until

is displayed.

bla

Release the button to

take a zero reading

(0.00 ppm).

Fill to 10 mL line

with sample. Cap

and mix until tablet

disintegrates. Make

readings within 30

seconds after

disintegration of the

tablet.

Add one Chlorine

DPD #3 IG Tablet

(6197A) to the sample

from Step 6. Crush

tablet.

READ