TROUBLESHOOTING

PROBLEM

CHECK

ACTION

Meter won’t turn on

Battery

Replace

AC Adapter

Plug in

AC Wall Outlet

Verify power source

Contact LaMotte for
Return Authorization

Return to LaMotte for
repair

Suspect Calibration

Check calibration with
standards

Use new standards

Verify standard
concentration

Run test with alternative
reagent system

Verify with another
meter

Check other meter
calibrations

Check tube alignment

Re-align tube

Check sample tubes for
dirt and scratches

Check, clean and/or
replace if necessary

Check to see if internal
meter components are
wet

Always dry tubes before
inserting. Examine
chamber for visible
moisture.

Reset meter to factory
calibration

With meter off, hold
down

t and press READ

Contact LaMotte for
Return Authorization

Return for calibration
check

Very low battery

Change battery

Over range

Dilute sample

Burnt out bulb

Call LaMotte

BAT

Low battery

Change battery