LaMotte 1200 Colorimeter User Manual
Page 55
Advertising
NOTE: If the molybdenum reading exceeds 50 ppm, repeat procedure on diluted sample,
and multiply the result by the appropriate dilution factor. See 1200 Colorimeter
Instruction Manual for procedure.
3
9.
10.
Wipe the tube dry.
Cap and mix
until the
powder
dissolves.
Solution will
turn yellow if
molybdenum is
present.
Align the index line
with the arrow on
the meter, insert
tube into chamber.
Close the lid. Push
the
button. Record
results as ppm
Molybdenum.
READ
CAL
I200
COLORIMETER
••••••••••••••••••
READ
18.0
11.
L Mott