MOLYBDENUM
TEST METHOD SPECIFICATIONS

APPLICATION

Boiler and cooling water

RANGE

0 to 30 ppm Molybdenum

METHOD

Calcium thioglycolate reacts with molybdenum to give a yellow color with an intensity
proportional to the amount of molybdenum present.

HANDLING & PRESERVATION

Molybdenum samples may be stored in either plastic or glass containers.

INTERFERENCES

Nickel levels less than 50 ppm do not interfere; aluminum levels less than 10 ppm do
not interfere; chromate, at higher concentrations, interferes due to the intense yellow
color. Ferrous iron levels below 50 ppm do not interfere, but low levels of ferric iron will
cause a large blank. Highly buffered samples may exceed the capacity of the system
possibly producing inaccurate results. Add an extra 1.0 mL of *Mo Buffer (3997) to
adjust the pH of the sample to approximately 4.5. Multiply the result in ppm by 1.08.