BROMINE TEST PROCEDURE - DPD METHOD
Read the 1200 Colorimeter Manual before proceeding. Carefully
wipe tubes dry before inserting into the colorimeter chamber.

FREE AVAILABLE BROMINE

Push the
button to turn the

meter on. Press the

button

and hold it for 2
seconds until

is

displayed. Release

the button to take a

zero reading
(0.00 ppm).

READ

ZERO

bla

Fill the Water

Sample Collecting

Bottle (0688) with

sample water. This

will be used to

dispense sample

water for the tests.

Insert the tube

into the

chamber, being

sure to align the

index line with

the arrow on

the meter.

Close the lid.

This tube is the

blank or zero.

Rinse and fill a

colorimeter tube

(0290) to the 10

mL line with

sample water.

Cap and wipe

dry.

Align the index line

with the arrow on

the meter, insert

tube into chamber.

Close the lid. Push

the

button. Record

results as ppm

Bromine.

READ

CAL

I200

COLORIMETER

••••••••••••••••••

READ

7.

5.

3.

1.

2.

4.

6.

I200

COLORIMETER

••••••••••••••••••

READ

ZERO

Empty all but a few

drops of sample from

tube. Add one

*Chlorine DPD #1

Tablet (6903A) and

crush with tablet

crusher (0175).

To insure

accurate results, tablet

must be crushed before

filling tube.

NOTE:

Fill to 10 mL line

with sample. Cap

and mix until tablet

disintegrates. Make

readings within 30

seconds after

disintegration of the

tablet.