MOLYBDENUM PROCEDURE - Thioglycolate Method

Read the 1200 Colorimeter Manual before proceeding. Carefully
wipe tubes dry before inserting into the colorimeter chamber

MOLYBDENUM

2

8.

Fill the Water

Sample Collecting

Bottle (0688) with

sample water. This

will be used to

dispense sample

water for the tests.

Rinse and fill a

colorimeter

tube (0290) to

the 10 mL line

with sample

water.

Use the 0.05 spoon

(0696) to add 0.05g

of *Molybdenum

Indicator Powder

(6486).

1.

2.

Insert the tube

into the

chamber, being

sure to align the

index line with

the arrow on

the meter.

Close the lid.

This tube is the

sample blank.

6.

5.

Wipe the tube dry.

Use the 1.0 mL

pipet (0354) to

add 1 mL of

*Molybdenum

Oxidizing

Reagent

(6485). Cap

and mix.

4.

Use the 1.0 mL

pipet (0372) to

add 1 mL of *Mo

Buffer (3997).

Cap and mix.

3.

7.

Push the
button to turn the

meter on. Press the

button

and hold it for 2
seconds until

is

displayed. Release

the button to take a

blank reading

(0.0 ppm).

READ

ZERO

bla

I200

COLORIMETER

••••••••••••••••••

READ

ZERO

bla

L Mott