analyze this sample since a pinkish-red solution absorbs mostly green light.

REAGENT BLANK

Some tests will provide greater accuracy if a reagent blank is determined to

compensate for any color or turbidity resulting from the reagents themselves. A

reagent blank is performed by running the test procedure on demineralized or

deionized water. Use sample water to SCAN BLANK. Insert the reacted reagent

blank in the colorimeter chamber and select SCAN SAMPLE. Note result of

reagent blank. Perform the tests on the sample water as described. Subtract

results of reagent blank from all subsequent test results. NOTE: Some tests

require a reagent blank to be used to SCAN BLANK.

COLORIMETER TUBES

Colorimeter tubes which have been scratched through excessive use should

be discarded and replaced with new ones. Dirty tubes should be cleaned on

both the inside and outside. Fingerprints on the exterior of the tubes can cause

excessive light scattering and result in errors. Handle the tubes carefully, making

sure the bottom half of the tube is not handled.

LaMotte Company makes every effort to provide high quality colorimeter tubes.

However, wall thicknesses and diameter of tubes may still vary slightly. This may

lead to slight variations in results (e.g. if a tube is turned while in the sample

chamber, the reading will likely change slightly). To eliminate this error put the

tubes into the sample chamber with the same orientation every time.

The tubes that are included with the colorimeter have an index mark to facilitate

this. If possible, use the same tube to SCAN BLANK and SCAN SAMPLE.

METER CARE

The optical system of the SMART3 must be kept clean and dry for optimal

performance. Dry the colorimeter tubes before placing them in the chamber to

avoid introducing moisture. For best results store the instrument in a area that is

dry and free from aggressive chemical vapors.