Procedure – LaMotte SMART 3 Electronic Soil Lab User Manual

Page 84

Advertising
background image

84

SMART3 Soil 1.11

PROCEDURE

Press and hold

1.

until colorimeter turns on.

Press

2.

ENTER

to select TESTING MENU.

Select

3.

ALL TESTS (or another sequence containing 089 Sulfate HR) from

TESTING MENU.
Scroll to and select

4.

089 Sulfate HR from menu.

Use the 1 mL pipet (0354) to add 1 mL of *Sulfate Extracting Solution (6363)

5.

to the test tube (0701) then add deionized water to the 15 mL line.
Add 3 one gram measures of soil using the 1 g spoon (0697). Cap vial and

6.

shake for five minutes.
Filter and collect all of the soil filtrate using the funnel (0459) and filter paper

7.

(0463). If the filtrate is not clear, filter a second time.
Fill a clean tube (0290) to the 10 mL line with the soil extract.

8.

Insert tube into chamber, close lid and select SCAN BLANK..

9.

Remove tube from colorimeter. Use the 0.1 g spoon (0699) to add one

10.

measure of *Sulfate Reagent (V-6277). Cap and shake until powder

dissolves. A white precipitate will develop if sulfates are present. Wait 5

minutes.
Mix tube again. Insert tube into chamber, close lid and select SCAN

11.

SAMPLE. Multiply the result by 1.65 to determine the sulfur concentration in

ppm.
Press

12.

to turn colorimeter off or press

EXIT

to exit to a previous menu

or make another menu selection.

SULFUR CONCENTRATION CHART

PARTS PER MILLION

RANGE

0-16 ppm

Low

17-30 ppm

Medium Low

31-50 ppm

Medium

52-75 ppm

High

Test P

rocedures