SMART3 Soil 1.11

PROCEDURE

Press and hold

1.

until colorimeter turns on.

Press

2.

ENTER

to select TESTING MENU.

Select

3.

ALL TESTS (or another sequence containing 051 Iron Bipyridyl)

from TESTING MENU.
Scroll to and select

4.

051 Iron Bipyridyl from menu.

Fill a clean tube (0290) to the 10 mL line with the soil extract then neutralize

5.

according to the procedure on page 50.
Insert tube into chamber, close lid and select SCAN BLANK.

6.

Remove tube from colorimeter. Use the 0.5 mL pipet (0353) to add one

7.

measure of *Iron Reagent #1 (V-4450). Cap and mix.
Use the 0.1 g spoon (0699) to add 0.1 g of *Iron Reagent #2 Powder

8.

(V-4451). Cap and shake vigorously for 30 seconds. Wait three minutes for

maximum color development.
At the end of 3 minute waiting period, do not mix. Insert tube into chamber,

9.

close lid and select SCAN SAMPLE. Multiply the result by 5 to determine the

iron concentration in ppm.
Press

10.

to turn colorimeter off or press

EXIT

to exit to a previous menu

or make another menu selection.

NOTE: For best possible results, a reagent blank should be determined

to account for any contribution to the test result by the reagent system. To

determine the reagent blank, follow the above test procedure to scan a distilled

or deionized water blank. Then follow the above procedure to perform the test

on a distilled or deionized water sample. This test result is the reagent blank.

Subtract the reagent blank from all subsequent test results of unknown samples.

It is necessary to determine the reagent blank only when a new lot number of

reagents is obtained.

IRON CONCENTRATION CHART

PARTS PER MILLION

RANGE

0.0-1.3 ppm

Very Low

1.4-3.0 ppm

Low

3.0-5.0 ppm

Medium

5.0-10.0 ppm

Medium High

Above 10.0-25.0 ppm

High