Procedure – LaMotte SMART 3 Electronic Soil Lab User Manual

Page 58

Advertising
background image

58

SMART3 Soil 1.11

PROCEDURE

Press and hold

1.

until colorimeter turns on.

Press

2.

ENTER

to select TESTING MENU.

Scroll to and select

3.

ALL TESTS (or another sequence containing 005

Ammonia-N HR) from TESTING MENU.
Scroll to and select

4.

005 Ammonia-N HR from menu.

Use the 1 mL pipet (0354) to transfer 2 mL of soil extract into a clean tube

5.

(0290). Dilute to the 10 mL line with deionized water. Mix and neutralize

according to the procedure on page 56.
Insert tube into chamber, close lid and select SCAN BLANK. (See Note)

6.

Remove tube from colorimeter. Add 12 drops of Ammonia Nitrogen Reagent

7.

#1 (V-4797). Cap and mix. Wait 1 minute.
Use the 1.0 mL pipet (0354) to add 1.0 mL of *Ammonia Nitrogen Reagent

8.

#2 (V-4798). Cap and mix. Allow 5 minutes for maximum color development.
At end of the 5 minute waiting period, immediately mix, insert tube into

9.

chamber, close lid and select SCAN SAMPLE. Multiply the result by 50 to

determine the ammonia-nitrogen concentration in lb/acre.
Press

10.

to turn the colorimeter off or press the

EXIT

exit to a previous

menu or make another menu selection.

NOTE: It is strongly suggested that a reagent blank be determined to account

for any contribution to the test result by the reagent system. To determine the

reagent blank, follow the above test procedure to scan a distilled or deionized

water blank. Then follow the above procedure to perform the test on a distilled

or deionized water sample. This test result is the reagent blank. Subtract

the reagent blank from all subsequent test results of unknown samples. It

is necessary to determine the reagent blank only when a new lot number of

reagents is obtained.

AMMONIA NITROGEN CONCENTRATION CHART

POUNDS PER ACRE

RANGE

0-24 lb/acre

Low

25-68 lb/acre

Medium

Over 71 lb/acre

High

Test P

rocedures

Advertising