SMART3 Soil 1.11

Follow the test procedure

6.

in the manual to test the
prepared standard. Insert the
blank into the chamber. Close
the lid. Press

ENTER

to scan

the blank. The screen will
display Blank Done for about
1 second and then return to
the Test Menu.

002 Aluminum

Scan Blank

Scan Sample

12:00:00 001/500

Insert the reacted standard

7.

solution into the chamber.
Close the lid. Press

ENTER

to

scan the sample. The screen
will display Reading for about
1 second. The result will
appear on the screen.

002 Aluminum

0.28

ppm

Scan Blank

Scan Sample

12:00:00 001/500

The displayed result can now

8.

be standardized. Press

or

to scroll to calibrate.

002 Aluminum

0.28

ppm

%T/Abs

Calibrate

12:00:00 001/500

Press

9.

ENTER

to select

Calibrate. A reverse font
(dark background with light
characters) will appear to
indicate that the reading can
be adjusted.

002 Aluminum

0.28

ppm

%T/Abs

Calibrate

12:00:00 001/500

Testing