LaMotte SMART 3 Electronic Soil Lab User Manual
Page 26
26
SMART3 Soil 1.11
Follow the test procedure
6.
in the manual to test the
prepared standard. Insert the
blank into the chamber. Close
the lid. Press
ENTER
to scan
the blank. The screen will
display Blank Done for about
1 second and then return to
the Test Menu.
002 Aluminum
Scan Blank
Scan Sample
12:00:00 001/500
Insert the reacted standard
7.
solution into the chamber.
Close the lid. Press
ENTER
to
scan the sample. The screen
will display Reading for about
1 second. The result will
appear on the screen.
002 Aluminum
0.28
ppm
Scan Blank
Scan Sample
12:00:00 001/500
The displayed result can now
8.
be standardized. Press
or
to scroll to calibrate.
002 Aluminum
0.28
ppm
%T/Abs
Calibrate
12:00:00 001/500
Press
9.
ENTER
to select
Calibrate. A reverse font
(dark background with light
characters) will appear to
indicate that the reading can
be adjusted.
002 Aluminum
0.28
ppm
%T/Abs
Calibrate
12:00:00 001/500
Testing