LaMotte SMART 3 Electronic Soil Lab User Manual
Page 23
SMART3 Soil 1.11
23
Insert the blank into the
6.
chamber. Close the lid.Press
ENTER
to scan the blank. The
screen wil display Blank
Done for about 1 second and
then return to the Test Menu.
002 Aluminum
Scan Blank
Scan Sample
12:00:00 001/500
Insert the reacted sample
7.
into the chamber. Close
the lid. Press
ENTER
to scan
the sample. The screen will
display READING for about 1
second. The result will appear
on the screen.
002 Aluminum
1.00
ppm
Scan Blank
Scan Sample
12:00:00 001/500
To repeat the test, press
8.
ENTER
to scan the sample again. The
last blank scaned is used by
the colorimeter for repeated
scans. A different blank can
be used by pressing
or
to scroll to Scan Blank
and then scanning another
blank. Scroll with
or
and make another
selection with
ENTER
. The %T
or Absorbance of the last test
can be viewed by choosing
%T/Abs. Press
EXIT
to
escape to previous menus.
NOTE: The menus loop in
this screen so either
or
will lead to the menu
selection needed.
002 Aluminum
1.00
ppm
Scan Bank
Scan Sample
12:00:00 001/500
Testing