LaMotte SMART 3 Electronic Soil Lab User Manual

Page 24

Advertising
background image

24

SMART3 Soil 1.11

CALIBRATING LaMOTTE PRE-PROGRAMMED TESTS

The LaMotte Pre-Programmed Tests have been pre-calibrated. Recalibration of

the pre-programmed tests by the user is not possible. However, a procedure

to standardize the calibration can be performed to obtain the most accurate

readings or to meet regulatory requirements.

The LaMotte Pre-Programmed tests are standardized with one standard

solution. To standardize over the full range of the test, the concentration of

the standard should be chosen from the high end of the range. Alternatively, if

samples do not cover the full range of the test, a standard should be chosen

that is close to the concentration of the samples.

For the SMART3 Soil colorimeter, the standard should be prepared in distilled

or deionized water for the range of the reagent system before the multiplication

factor has been applied to the reading on the display. The following standards

are recommended to standardize over the full range of the tests:

Ammonia Nitrogen 3.00 ppm Ammonia Nitrogen
Copper

4.00 ppm Copper

Iron

4.00 ppm Iron

Manganese

11.00 ppm Manganese

Nitrate Nitrogen

2.00 ppm Nitrate Nitrogen

Nitrite Nitrogen

0.60 ppm Nitrite Nitrogen

Phosphorus

2.00 ppm Phosphate

Potassium

7.00 ppm Potassium

Sulfur

75 ppm Sulfate

Zinc

2.00 ppm Zinc

The standardization procedure should be followed as often as required by

regulations and laws for compliance monitoring.

In the example below, the Aluminum calibration will be standardized.

Prepare a standard solution to be tested. In this example, 0.30 ppm aluminum.

Testing