Procedure – LaMotte SMART 3 Electronic Soil Lab User Manual

Page 70

Advertising
background image

70

SMART3 Soil 1.11

PROCEDURE

Press and hold

1.

until colorimeter turns on.

Press

2.

ENTER

to select TESTING MENU.

Select

3.

ALL TESTS (or another sequence containing 056 Manganese HR)

from TESTING MENU.
Scroll to and select

4.

056 Manganese HR from menu.

Fill a clean tube (0290) to the 10 mL line with the soil extract then neutralize

5.

according to the procedure on page 50.
Insert tube into chamber, close lid and select SCAN BLANK.

6.

Remove tube from colorimeter. Use the 0.1 g spoon (0699) to add two

7.

measures of Manganese Buffer Reagent (6310). Cap and mix until powder

dissolves.
Use the 0.15 g spoon (0727) to add one measure of *Manganese Periodate

8.

Reagent (6311). Cap and shake for one minute. An undissolved portion of

the reagent may remain in the bottom of the tube without adversely affecting

the test results. Wait two minutes for maximum color development. Solution

will turn pink if manganese is present.
At the end of the two minute waiting period, mix, insert tube into chamber,

9.

close lid and select SCAN SAMPLE. Multiply the result by 5 to determine the

manganese concentration in ppm.
Press

10.

to turn colorimeter off or press

EXIT

to exit to a previous menu

or make another menu selection.

MANGANESE CONCENTRATION CHART

PARTS PER MILLION

RANGE

0-5 ppm

Low

5-12 ppm

Medium

13-24 ppm

Medium High

25-40 ppm

High

Over 40 ppm

Very High

Test P

rocedures