Battery – LaMotte SMART 3 Electronic Soil Lab User Manual

Page 47

Advertising
background image

SMART3 Soil 1.11

47

BATTERY

BATTERY/AC OPERATION

The SMART3 may be operated on battery power or using a USB wall adapter or

USB computer connection. If using the meter as a bench top unit, use the wall

adapter if possible to extend the battery life. The meter will remain on when the

USB adapter is used.

To charge the battery with the wall adapter, plug the smaller end of the USB

cable (USB mini B connector) into the meter and the larger end of the USB

cable (USB Type A connector) into the wall adapter. Plug the wall adapter into

an AC outlet. Reinsert the USB port plug after charging.

To charge the battery from a computer, plug the smaller end of the USB cable

(USB mini B connector) into the meter and the larger end of the USB cable (USB

Type A connector) into a USB port on a computer. Reinsert the USB port plug

after charging.

The battery icon will show no bars and flash when the unit first turns on. Then

the indicator will indicate the battery status by showing 0, 1, 2, 3 or 4 bars.

It will take 5 hours to fully charge a low battery. The battery icon will flash when

the battery is charging. The battery icon will show four bars and stop flashing

when it is fully charged. The charging circuit will automatically switch to a float

charge when the battery is fully charged. The charger may remain connected.

Some computers will NOT supply power to their USB ports during standby

operation. The wall adapter will charge the unit continuously.

The battery icon will show no bars and continuously flash if the battery is getting

low but the unit will still operate normally. A “Low Battery” message on the

status bar of the display will replace the time when the battery voltage is too

low for proper operation and accuracy may be degraded. A “Shutdown Low

Batt” message on the display will appear for a few seconds before the power is

switched off when the battery is too low to operate the unit.

To extend the battery life:

Shut down the unit with the power switch when not taking measurements

or use the power save option to have the unit automatically turn off after 5

minutes.
Store the unit in a cool dry place.

Fully charge the battery before storing the unit for extended periods of time.

Limit backlight use. The unit consumes 3X normal power with the backlight

on. Set the backlight time option to 10 seconds, or select “Button Control”

and keep the backlight off.

BATTERY REPLACEMENT


The lithium ion battery used in this unit should last for many years with

normal use. When it no longer powers the unit long enough to meet testing