Procedure – LaMotte SMART 3 Electronic Soil Lab User Manual

Page 64

Advertising
background image

64

SMART3 Soil 1.11

PROCEDURE

Carefully read the LaMotte Direct Reading Titrator Manual (1649) before

performing the titration procedure described below. The Titrator is calibrated in

terms of parts per million chloride and each minor division on the Titrator scale

equals 20 ppm.

Fill a clean test tube (0701) to the 15 mL line with deionized water.

1.

Add 3 measures of soil using the 1 g spoon (0697). Cap tube and shake for

2.

five minutes.
Filter and collect all of the soil filtrate using the funnel (0459) and filter paper

3.

(0463). The extract does not have to be clear since a slight turbidity does

not interfere in the test.
Fill the test tube (0778) to the 10 mL line with the filtrate.

4.

Add three drops of *Chloride Reagent #1 (4504) to the sample. Cap and

5.

shake to mix. A yellow color will result.
Fill the Direct Reading Titrator (0384) with *Silver Nitrate, 0.141 (3062DR) in

6.

the manner described in the instruction manual.
Titrate the test sample with *Silver Nitrate, 0.141 (3062DR) until the yellow

7.

color changes permanently to pink. Read the test result from the scale

where the large ring on the Titrator meets the Titrator barrel. If the plunger

reaches the bottom mark (1000 ppm) on the Titrator scale before the

endpoint color change occurs, refill the Titrator and continue the titration

procedure. Be sure to include the value of the original amount added (1000

ppm) when recording the final result.

Test P

rocedures

Advertising