Section 6 - Communication

I.

Page 94

1/06/2012 REV 1.10.2

The following tables show the mapping of the recorder parameters into Modbus registers. Knowledge of these
Modbus register assignments is only needed to make use of third party communications software. Both
registers

– high and low, must be accessed together to ensure valid data. The data is Modbus default byte

order and the registers DO NOT use zero byte addressing. Data encoding is first word low (32 bit data).

Table 6.3.2.2 Holding Registers 4xxx (Use Function 3 to Access)

32 Bits Modicon

TM

Format Floating Point

Writing to these registers has no effect

Register #

FUNCTION AND STATE

Register #

FUNCTION AND STATE

4361

Point 1 Data Low Word

4379

Point 10 Data Low Word

4362

Point 1 Data High Word

4380

Point 10 Data High Word

4363

Point 2 Data Low Word

4381

Point 11 Data Low Word

4364

Point 2 Data High Word

4382

Point 11 Data High Word

4365

Point 3 Data Low Word

4383

Point 12 Data Low Word

4366

Point 3 Data High Word

4384

Point 12 Data High Word

4367

Point 4 Data Low Word

4385

Point 13 Data Low Word

4368

Point 4 Data High Word

4386

Point 13 Data High Word

4369

Point 5 Data Low Word

4387

Point 14 Data Low Word

4370

Point 5 Data High Word

4388

Point 14 Data High Word

4371

Point 6 Data Low Word

4389

Point 15 Data Low Word

4372

Point 6 Data High Word

4390

Point 15 Data High Word

4373

Point 7 Data Low Word

4391

Point 16 Data Low Word

4374

Point 7 Data High Word

4392

Point 16 Data High Word

4375

Point 8 Data Low Word

4393

Point 17 Data Low Word

4376

Point 8 Data High Word

4394

Point 17 Data High Word

4377

Point 9 Data Low Word

4395

Point 18 Data Low Word

4378

Point 9 Data High Word

4396

Point 18 Data High Word

16 bit Integer registers accessible by Modbus RTU Serial only are available

– contact factory for detail.