11 registry, 12 languages, 13 system reflash – Monarch Instrument DC6000 User Manual

Page 94: Registry, Languages, System reflash, Language, Details here

Advertising
background image

Section 5

– Recorder Setup

I.

Page 86

1/06/2012 REV 1.10.2

Figure 5.7.11

– Language Screen

5.7.11 Registry

The Registry option allows the user to clear or save the registry. The Registry holds system base
setup values not in the configuration file and needs to be saved if system setup changes have
been made

– for example recalibrating the stylus, or may need to be cleared if upgrading the

firmware. This feature is entered from the

system admin menu

by pressing the icon shown right.

This will bring up the dialog box shown on the left. The user has three choices. The top button
will save the registry settings to system nonvolatile memory. The middle button will clear the
registry and set it to factory default values

Note: Do not clear the registry unless required

to do so by a

system firmware upgrade

. The lower button allows the user to exit without

affecting the Registry.

5.7.12 Languages

The languages option allows the users to customize the unit language for the local country by
loading preconfigured language files. This
feature is entered from the

system admin menu

by pressing the icon shown right.

This will bring up the language dialog screen. NOTE:
Loading a new language will unload the existing
configuration file
.

Select the pull down menu to display a list of available
languages. The recorder will check the internal memory for
the language file. The user needs to load the language file
“MultiLanguage.CSV” onto the root of the internal memory
card. This is a spreadsheet style file with English in the first
column and alternative languages in subsequent columns. It
is (and needs to be) in Comma Separated Variable format.

This language file can be edited using a standard spreadsheet editor such as Excel

TM

and any language can

be added following a simple set of rules

– See Appendix C

Note: Save the file as .CSV and NOT .XLS format.

Select the language of choice and press the okay button.

The recorder will change to the new

language on the fly. The default language is English.

5.7.13 System ReFlash

This option allows the administrator to upgrade the system firmware. Once started this operation
cannot be reversed. Choosing to proceed will require a system reset even if you cancel the
operation.

DO NOT RESET the unit, DO NOT REMOVE POWER and DO NOT PRESS THE ABORT BUTTON while
this operation is in process as it WILL MAKE THE UNIT INOPERABLE.

Before proceeding you need the image file that will be loaded. This file is called NK.bin and is
available from your service center or can be downloaded from the web. Ensure that you need to
update the firmware; often firmware upgrades will involve clearing the

Registry

,

new

configuration

and

language files

. Read the text file that accompanies any

firmware upgrade release. Load the file (NK.bin) onto removable media

– compact

flash drive or USB thumb drive, and insert this into the recorder before starting this
procedure.

Select the reflash icon shown left from the

System Admin Menu

. Read and accept the

warning notice shown right and proceed with caution.

You then need to accept or skip erasing the registry

– this information will be in the release notes with the

firmware. In some instances, failure to erase the registry will cause the unit to malfunction after reflash.