2 environment and location, 3 installation, Environment and location – Monarch Instrument DC6000 User Manual

Page 23: Installation

Advertising
background image

Section 3

– Installation

I.

Page 15

1/06/2012 REV 1.10.2

3.2 Environment and Location

The instrument is intended to operate in the following environment:

Indoor Use Only

Installation Category II

per IEC 664

Pollution Degree Level II

per IEC61010-1

Measurement Category I

per IEC61010-1

Altitude

up to 2,000 m

Temperature

-10 °C to +50 °C operating per IEC61010-1

Humidity

Maximum relative humidity 80% for temperatures up to
31°C decreasing linearly to 50% relative humidity at 40°C

AC Mains Supply

100 - 240 Vac ~ +10% 50/60 Hz 35 VA

NOTE:

The recorder is designed to be panel mounted and as such should be considered as fixed
equipment or permanently connected. If permanently connected, disconnection from the supply
must be possible via a customer supplied switch or circuit breaker rated at 120V or 240V
(dependent on local voltage supply) 5A minimum when connected to an AC supply or 30V 2A
minimum when connected to a DC supply. This disconnection device must disconnect all
current-carrying conductors. It must be included in the panel installation and should be clearly
marked, in close proximity to the Recorder and easily accessible to the operator.

3.3 Installation

The recorder is sized to fit in a standard DIN panel cutout of 5.43” x 5.43” [138 mm x 138 mm] and requires a
minimum of 6.5” [188.2 mm] behind panel depth not including power and input source cable space needed.
Actual dimensions are shown below.

Figure 3-1 Recorder Dimensions in Inches [mm]

WARNING: Do not use this instrument in any manner inconsistent with these operating

instructions or under any conditions that exceed the environmental specifications stated.

Advertising